53
Practical Example
Determining an Unknown Piece Count; Weighing in the Preset Reference Sample Quantity
Settings (changes in the factory settings required for this example):
Setup: App: Application 1: Counting: Average piece weight updating: Manual
Setup: Printout: Application-defined output: Auto print upon initialization: All values
Step Key (or instruction) Display/Output
1. Delete previous setting if necessary c
2. Prepare a container for the Place the empty container
parts to be counted on the balance
3. Tare the balance w
4. Place the reference sample quantity Place the displayed number
on the balance (example: 10 pcs, of parts in the container
each weighing 2.148 g)
5. Determine the average piece weight
Start soft key
(number of decimal places displayed
depends on the balance model)
6. If necessary, increase the number of parts Place additional parts
and perform reference sample updating in the container
(example: 7 additional pieces)
Opt. soft key
7. Weigh uncounted parts Place parts to be counted
in the container
8. If desired, print total piece count r
(here: 153 pcs)