Practical Example
Displacement: Determine the Density of Samples of a Solid Using the Displacement Method.
Reference Liquid: Ethylene glycol. Temperature: 23°C (73.4 °F). Set the density at 20°C (68°F) to 1.113 g/cm
2
and the volume expansion coef-
ficient at 20°C (68°F) to 0.00064 grd
–1
.
Settings (changes in the factory settings required for this example):
Setup: App: Application 1: Density: Method: Displacement
Setup: App: Application 1: Density: Liquid for buoyancy: User-definable
Step Key (or instruction) Display/Output
1. Delete previously stored values c
if necessary
2. Change configurations:
Param. soft key
Temperature: 23.0
Density of ref. liquid: 1.113
Expansion coefficient: 0.000640
3. Place the container with the
reference liquid on the balance
4. Tare the balance w
5. Determine the weight of the
sample in air: place sample
on the weighing pan
6. Store weight value
Wa soft key
7. Determine the weight of the sample in liquid:
place sample in the sample holder
8. Store weight
Wfl soft key
Density of sample is displayed
9. Display volume of sample
Vol. soft key
10. Display weight
Weigh soft key
11. Repeat procedure with next sample, if desired
Start soft key