95
Practical Example
Totalize counted pieces and print out statistics
Settings (changes in the factory settings required for this example):
Setup: App: Application 1: Counting: Average piece weight updating: Manual
Setup: App: Application 3: Statistics: Evaluated values: Calculated
Setup: App: Application 3: Statistics: Evaluation mode, MR function: Final evaluation, display + print
Step Key (or instruction) Display/Output
1. Turn on the balance and configure e
the settings as indicated above
2. Delete old statistics data, c
if necessary
3. Tare the balance w
4. Toggle to Application 1: Counting d
5. Place the displayed number of parts Place parts to be counted
on the balance (here: 10 pcs) on the balance
6. Initialize the Counting application
Start soft key
7. Remove the reference sample Unload the balance
quantity and toggle to Statistics d
8. Place a number of parts on the Place parts on the balance
balance (here: 35 pcs)