Settings
26
 Setup
    wp-1 
1
      1.9.      Calibration and Adjustment
        1.9.1 *    Ext. calibration/adjustment; default weight
        1.9.3    External calibration/adjustment; weight can be selected under menu item 1.18.1
        1.9.4    Internal calibration/adjustment (models with built-in motorized calibration weight only)
        1.9.8    Setting the preload
        1.9.9    Deleting the preload
        1.9.10    No function when you press the 
J key
      1.10.       Calibration/Adjustment sequence
        1.10.1    Calibration with automatic adjustment
        1.10.2 *   Calibration with manual adjustment
      1.11.       Zeroing range
        1.11.1    1 percent/max. cap.
        1.11.2 *   2 percent/max. cap.
      1.12.       Initial zero point range
        1.12.1 *   Factory setting (depending on the model type)
        1.12.2    2 percent/max. cap.
        1.12.3 *   5 percent/max. cap.
      1.13.       Tare/zero at power on:
        1.13.1 *   On
        1.13.2    Off, load previous tare values
        1.133    Only zero at power on
      1.15.       Calibration prompt for connected IS weighing platform
        1.15.1 *   Off
        1.15.2    Calibration prompt W flashes on the display
      1.16.       External calibration 
1
)
        1.16.1 *   Activated
        1.16.2 
2
)  Blocked
      1.17.       Calibration weight unit 
3
)
        1.17.1 *   Grams
        1.17.2    Kilogram
        1.17.4 
1
)   Pounds
      1.18.       Entering calibration weight
        1.18.1    External user calibration weight (enter value, e.g.: 10,000 kg) 
      3. 1.       Weight unit 2 
3
)
        3.1.1    Grams /o 
1
)
        3.1.2    Grams /g
        3.1.3 *    Kilograms /kg
        3.1.4    Carats /ct 
1
)
        3.1.5    Pounds /lb 
1
)
        3.1.6    Ounces /oz 
1
)
        3.1.7    Troy ounces /ozt 
1
)
        3.1.8    Hong Kong taels /tlh 
1
)
        3.1.9    Singapore taels /tls 
1
)
        3.1.10    Taiwan taels /tlt 
1
)
        3.1.11    Grains /GN 
1
)
        3.1.12    Pennyweights /dwt 
1
)
        3.1.14    Parts per pound /lb 
1
)
        3.1.15    Chinese taels /tlc 
1
)
        3.1.16    Mommes /mom 
1
)
        3.1.17    Austrian carats /K 
1
)
        3.1.18    Tola /tol 
1
)
        3.1.19    Baht /bat 
1
)
        3.1.20    Mesghal /MS 
1
)
        3.1.21    Tons /t 
1
)
        3.1.22    Pound: Ounces 
1
)
      3.2.      Display accuracy 2
1
)
        3.2.1 *    All digits
        3.2.2    Reduced by 1 decimal place for load change
        3.2.14    10-fold increased resolution
        3.2.15    Increase resolution by 2 scale intervals
        3.2.16    Increase resolution by 1 scale interval
      3.3.      Weight unit 3
3
): Parameters, see 3.1 Weight unit 2
      3.4.      Display accuracy 3
1
): Parameters, see 3.2 Weight unit 2
      9.1.      Restore factory settings in WP1 numeric menu
        9.1.1    Yes 
        9.1.2 *    No
1
) = Not available on instruments verified for use in legal metrology 
2
) = Factory setting on instrument verified for use in legal metrology 
3
) = Menu depends on weighing platform model