Page 37
Art. no.: 2.511076 Rev.: g
BP recording 3
User Guide Long Term Recording 3.4
BR-102 plus / BR-102 plus PWA
9. Check that the initial measurement is successful.
10. After checking the first measurement and battery display, position
the BR-102 plus / BR-102 plus PWA in the pouch and secure.
– Subsequent measurements are taken as defined for the program
selected.
– During BP measurement, the ascending/descending cuff pres-
sure is shown. After a measurement has been taken, the result is
displayed for approximately one minute (see Displaying a Re-
cording on the BR-102 plus, page 39).
Initial Measurement Check
After the first measurement (on the resting patient), it is recommended
that the following is checked to help ensure that the recording is
successful:
• Ensure that a BP reading has been obtained and is displayed on the
BR-102 plus / BR-102 plus PWA.
• Check that the measurement has been taken using the auscultatory
method (KOR displayed next to the measurement, below the time).
Note: If the cuff does not have a microphone (or an unsuccessful
auscultatory measurement has been taken) OSC is displayed.
• If the measurement has not been taken correctly check the cuff
placement and microphone position.
• Check the battery display and ensure that it still shows full capacity.
12.02.2014 10:19
KOR
SYS 128 99 mmHg
mmHg
DIA 86 71 bpm
OK
PUL
MA