EasyManua.ls Logo

Schiller BR-102 PLUS - Safety Conscious Operation; Safety Facilities

Schiller BR-102 PLUS
76 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Page 9
Art. no.: 2.511076 Rev.: g
Safety Notes 1
User Guide Maintenance 1.6
BR-102 plus / BR-102 plus PWA
1.6 Maintenance
1.7 Hygiene
1.8 Safety Conscious Operation
1.9 Safety Facilities
There are no serviceable parts inside. Refer servicing to qualified
technician authorised by SCHILLER only.
Do not use high temperature sterilisation processes (such as
autoclaving). Do not use E-beam or gamma radiation sterilisation.
Do not use solvent or abrasive cleaners on either the unit or cable
assemblies.
Do not, under any circumstances, immerse the device or cable
assemblies in liquid.
For cleaning and disinfection observe the legal requirements
applicable.
Only use cleaning agents and disinfectants recommended by
SCHILLER. Unsuitable agents can damage the device. Clean and
disinfect the device in accordance with the instructions given in
this book.
Make sure that the staff have read and understood the operating
instructions - particularly this Safety Notes section.
Immediately report any changes that impair safety (including
operating behaviour) to the person responsible.
Only use accessories and other parts recommended or supplied
by SCHILLER AG. Use of other than recommended or supplied
parts may result in injury, inaccurate information and/or damage to
the recorder.
Connecting the unit to a PC with defective cables may constitute a
danger to life. Therefore:
Do not connect the BR-102 plus / BR-102 plus PWA to any PC if
the earth connection is suspect or if the mains lead is damaged
or suspected of being damaged.
Damaged cable connections and connectors must be replaced
immediately.

Table of Contents

Related product manuals