Page 53
Art. no.: 2.511076 Rev.: g
Maintenance 6
User Guide Visual Inspection 6.1
BR-102 plus / BR-102 plus PWA
6 Maintenance
6.1 Visual Inspection
Visually inspect the unit, cables, connectors, tubing and cuff for the
following:
Device casing not broken or cracked.
Display not broken or cracked.
Microphone cable sheathing and connectors not damaged, no
kinks.
No kinks, abrasion or wear in the tubing assembly.
Pressure bladder and tube connector in good condition.
No excessive soiling or damage of the cuff and velcro fastening.
All maintenance work must be carried out by a qualified technician
authorised by SCHILLER AG. Only maintenance procedures given in
this book may be carried out by the user.
The following table indicates the maintenance intervals, the
maintenance requirement, and the person authorised to carry out the
procedure.
Interval Maintenance Responsible
Every 6 months Visual inspection of the monitor, cables
and tubes and cuff (see below).
User
Every 24 months Measurement calibration. SCHILLER AG authorised
service centre.
Do not use if the unit, or any cable assembly or accessory, is
damaged.
Defective units, damaged cables, or damaged accessories must
be replaced immediately.