EasyManua.ls Logo

Schiller BR-102 PLUS - Safety Standards

Schiller BR-102 PLUS
76 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Page 61
Art. no.: 2.511076 Rev.: g
Technical Data 7
User Guide Error Messages 6.5
BR-102 plus / BR-102 plus PWA
Safety Standards IEC 60601-1
ISO 81060-2:2013
a
IEC 80601-2-30
IEC 60601-1-2 (EMC)
Protection Class Internal power supply
Protection Class of Casing IP42 when using the carrying pouch
Applied Part BF according to IEC/EN 60601-1
Conformity CE 0123 according to Annex II 93/42/EEC (medical devices)
Classification IIa according to MDD 93/42/EEC
Trusted Accuracy The BR-102 plus / BR-102 plus PWA is clinically validated to all
internationally recognised organisations:
BHS (in progress)
ESH (2002)
AAMI SP10:2002
Environmental conditions
(Operation)
Operating temperature
Relative humidity
Pressure during operation
0°C to + 40°C (+ 32°F to + 104°F)
15 % to 95 % (non-condensing)
540 hPa to 1060 hPa
Environmental Conditions
(storage and transport)
Transport temperature
Storage temperature
Relative humidity (storage
and transport)
Pressure (storage and
transport)
- 10°C to + 50°C (+ 14°F to + 122°F)
+ 5°C to + 50°C (+ 41°F to + 122°F)
10 % to 95 % (non-condensing)
500 hPa to 1060 hPa
a. The sphygmonometer was clinically investigated according to the requirements of ISO 81060-2:2013