INCU-SHAKE Compact Stackable shaking Incubator Manual V2023.1
9
4.2.4 Shaking Control
The direct beltless driving shaking system ensures smooth and reliable orbital
shaking movement with a speed range between 30-300rpm. The orbital diameter is
adjustable from 1mm to 50mm continuously. The loading weights must be below the
maximum rated, otherwise, it will prevent the shaking tray from smooth movement,
and eventually increase the wear out the bearings and joints.
4.2.5 Temperature Control
The heaters are behind the back wall. The temperature is controlled with a PT100
temperature sensor. The heat generated is distributed via cross flow fan evenly
across the inner working chamber.
4.2.6 Control Electronics
The microprocessor controller at the front panel is designed to service the precise
control requirements of the working chamber environment, providing optimum
programmable conditions for culture growth. The microcomputer is with 7’’ large touch
screen to permit efficient operator entry and settings of data.
Each stack of unit is operated independently with its own controller.
4.2.7 Shaking Tray Operation
The shaking movement of the tray is interrupted as soon as the door is opened by
more than 20° of angle. In the same time, both the circulating fan and heater stops
working at the same time when door is opened.
To unlock, turn the handle of the locking device to the left.
To lock, turn the handle of the locking device as far as possible to the right.
Shaking movement, circulating fan, and heating controlling resumes once the door is
closed.