INCU-SHAKE Compact Stackable shaking Incubator Manual V2023.1
38
6.12.2 Uninstalling the Sticky Mat
Release the Sticky Mat from one side of the tray and then pull upwards at an angle
until it’s completely detached from the shaking platform.
6.12.3 Adhesive Strength
The Sticky Mat will lose the adhesiveness temporarily when in contact with water,
but the full strength of adhesiveness will come back once it becomes completely
dry.
Clean up any spills as fast as possible. If water spills comes in contact with the
Sticky Mat, do not start operation until it's dried completely.
To ensure the full adhesive strength, the recommended temperature range of the
Sticky Mat is 18-45°C
Ensure the chamber is free of condensation during humidification. The vessels may
become unstuck from the mat, if there is condensation inside the incubation
chamber.
6.12.4 Safety
To ensure the safety of use of Sticky Mat, only use flasks and vessels of high
quality, for example, Schott Duran® Borosilicate glass, Corning® Polycarbonate
vessels. The use of all other materials or quality levels is in the responsibility of the
user.
It is recommended that test the Sticky Mat with each type of vessels to be used,
under the same condition (Speed, temperature, etc ) for a short time period. Do not
use it for this application If any movements or dislodgements are detected.
During high speed operations, the Sticky Mat is less secure than the flask clamps.
Refer to the below table for the maximum speed to use with a combination of
shaking diameter and flasks.