INCU-SHAKE Compact Stackable shaking Incubator Manual V2023.1
25
6.4.2 Operating Under Programmable Control Mode
This incubator can operate according to stored program to meet various requirements
of professional experiments. Under programmable control mode, it can be
programmed to save up to 10 different segments, each with their own Ramp Time,
Ramp End Temperature, Hold Time and Speed. At the end of the settings, set the
Hold Deviation and Cycle Times to complete the settings.
To set the general parameters under "Programmable" control mode, follow the below
steps,
⚫ Press "Stop" button in the home page to stop the current operation.
⚫ Switch the "CTRL MODE" in the "SET" page to "PROG" mode. Press "SET" in
the home page
⚫ Press "PROG" in the "SET" page to start setting the general parameters
(Excluding Photosynthetic Lighting)
SEG—Segment No.
The function indicates the serial number of current segment. No setting required.