INCU-SHAKE Compact Stackable shaking Incubator Manual V2023.1
28
"Shaking Flask" icon will start swing. The shaking motion of the shaking tray
can be observed through the front window. The temperature control starts from
the "Segment 1". Time starts counting from the "Ramp Time" of "Segment 1".
Pause the operation: During the operation, the incubator can be paused by
pressing the "STOP" button. Once it pauses, the "RUN" button keeps twinkling,
the shaking tray stops shaking motion, the timer pauses counting. The
temperature remains under control at the setpoint at the time of the pause
operation. Press the "RUN" to resume the current operation under programmable
mode.
Stop the operation: When the operation is paused, press the "STOP" button
once again, it will exit the current program operation. Once it's stopped, the
"Shaking Flask" stops, the "RUN" button dims, the shaking tray stops shaking
motion, the timer stops and will be reset.
The temperature controlling will also be switched off. However, if the "KEEP" in
the "FUNC" page is switched on, the temperature will remain under control at
their setpoints of last segment even when the program operation is stopped.
6.5 Operating Photosynthetic Lighting System (Option)
The photosynthetic lighting system features LED lighting panel with specific light
spectrums that is most productive for algae growth. The standard light panel
generate blue and red lighting. White LED light option provides similar lighting
spectrum like daylight.
Each color of the light can be controlled under fixed value mode, or programmable
mode, which runs day/night cycles.
Each color light intensity can be set with the range of 0-100% power output.
The lighting panel is protected by acrylic housing, which makes it easy to clean and
water proof.
6.5.1 Operating Under Fixed Value Control Mode
There is an individual set page for photosynthetic lighting system, which include
settings under both Fixed Value and Programmable control mode.
****