EasyManua.ls Logo

SciQuip Incu-Shake Compact - Technical Data; Unit Dimensions and Views

SciQuip Incu-Shake Compact
49 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
INCU-SHAKE Compact Stackable shaking Incubator Manual V2023.1
6
3. ENVIRONMENTAL REQUIREMENTS
Best working temperature shall be 18°C--30°C, and humidity no more than 80%RH.
In consideration of ventilation, heat dissipation and other factors, a minimum of 20 cm
space shall be reserved in the surrounding of the equipment during installation to
facilitate adequate convection of air around the unit.
Indoor use only. Do not locate the unit exposed to direct sunlight or near heating
/cooling ducts.
The incubator shall be placed on a flat, solid and fire-retardant platform which can
support its weight.
The unit must be kept away from electromagnetic interference sources.
There shall be no inflammable, explosive, corrosive, toxic or harmful gas that may
cause sharp temperature change in the vicinity of the equipment.
SciQuip only commits that the incubator can operate normally in the place where the
elevation is less than 2000m.
The power supply voltage of this equipment shall be allowed to change within ±10% of
the rated working voltage.
The transient over-voltage of the power supply shall be Class II.
The pollution degree is Class II. Please keep the chamber clean. Regular cleaning is
required.

Table of Contents