INCU-SHAKE Compact Stackable shaking Incubator Manual V2023.1
29
To set the lighting outputs of under "Fixed Value" control mode,
Press "SET" in the home page
Press "LIGHTS" in the "SET" page to start setting the photosynthetic lighting
system
Press the switch to switch the control mode to "FIXED"
LIGHT-RED Red Light Output button
Press it and use the keyboard to enter the red light output set value, and press OK
to confirm and save. Set range 0-100%.
LIGHT-BLUE Blue Light Output button
Press it and use the keyboard to enter the blue light output set value, and press OK
to confirm and save. Set range 0-100%.
Switch the "LIGHT MODE" in the "SET" page to "FIXED " mode.
Press "LIGHT" switch in the home page to switch "ON" the lighting system.
Press the switch again to switch "OFF" the lighting system.
6.5.2 Operating Under Programmable Control Mode
With the "Programmable" control mode, the photosynthetic lighting system can
simulate and run day/night cycles. Together with the other operating parameters, the
incubator can create reproducible conditions of "natural" environment.