INCU-SHAKE Compact Stackable shaking Incubator Manual V2023.1
33
DEF.PERIOD Defrosting Time
This function indicates the time duration for each single defrost cycle both under auto
and manual operations. Use the keyboard to enter the set value between 1-20 mins
and press OK to confirm and save. Set it as 0 to disable the defrosting function.
When the ambient temperature is higher than 25, like in summer, the recommended
setting is 2-3 minutes. As ambient conditions may vary, therefore, always try to find
the most appropriate settings for specified ambient condition.
6.7 Setting Clock Time
There is real clock time display in the right below corner of the homepage. To set the
clock time,
Attention:
It is considered as normal, that the actual temperature slightly
deviates during defrosting, and will be stabilized shortly after
defrosting cycle finishes.
.
****