INCU-SHAKE Compact Stackable shaking Incubator Manual V2023.1
19
Press it to enable or disable the photosynthetic LED lighting function.
Red Light Indicator:
The indicator lights up when the red light is on, with display of current
lighting output value in percentage.
Blue Light Indicator:
The indicator lights up when the blue light is on, with display of current
lighting output value in percentage.
Control Mode Status:
This box indicates the status of the current control mode, fixed value
mode or programmable mode.
If it is controlled under Fixed Value mode, no display.
If it is controlled under Programmable mode, it displays the current
segment number and remaining cycle times.
Speed Display:
It displays the current operating speed in rpm.
“Temperature” Display:
It displays the current operating temperature in °C.
Time Display:
It displays current accumulative running time in minutes.
“SV” Column Display:
It displays the current set points of operating parameters, Temperature,
Speed, and Timer.
Special Remind!
Under programmable mode, the SEG box shows as followings,
Pm/L’’n,
m: Current segment No.
n: Remaining cycle times.
Special Remind!
Under Fixed Value mode, the Time box in the “SV” column shows the
“Timer” setpoint;
Under programmable mode, the Time box in the “SV” column shows
the total time of “Ramp Time” + “Hold Time” of the current segment.