20
-
【▲】/【▼】
Chapter 6. Setting Recordeing Range
6-1. Setting temperature recording range
The temperature recording range
can be set by changing over the DIP
switches.
CAUTION
Before changing over the DIP
switch, be sure to turn OFF (
【 】
)
the power switch.
Setting range 0-50°C (30-130°F)
①
Turn OFF (
【 】
) the power switch, then turn ON the DIP
switch No. 3, and turn OFF the No. 4.
②
When the power switch is turned ON, temperature will be
recorded in the range from 0 to 50°C or 30 to 130°F.
Setting recording range manually
①
Turn OFF (
【 】
) the power switch, turn ON the DIP
switches No. 3 and No. 4, and then turn the power switch
ON again.
②
Press the
【
FUNCTION
】
key five times (Six times when
the SD card is inserted and the [SD Card Record] switch
is set to OFF) while the normal measurement screen is
displayed.
『
MODE
』
and
『
r1
』
will appear and the present
central temperature of the chart will blink on the left side.
『
- -
』
will blink instead if the recording range has not been
changed from the initial setting. (-20 to 80 °C or 0 to 170
°F)
③
Change the temperature to be recorded in the center of
the chart paper with the
【▲】
or
【▼】
key, then press the
【
ENTER
】
key. The central temperature will stop blinking
to show the setting has been completed.
④
Then, the present recording range will blink on the right
side of the screen.
⑤
Set the recording range 10 (+ / -10°C) or 25 (+ / -25°C)
in Celsius or 25 (+ / -25°F) or 50 (+ / -50°F) in Fahrenheit
with the
【▲】
or
【▼】
key, then press the
【
ENTER
】
key.
The recording range will stop blinking to show that set-
ting has been completed.
Reference
If this setting is not required, press the
【FUNCTION】
key twice to return to the normal display.
ON
3 4
ON
3 4
ON
3 4
Sett ing Temperature
Record ing Range w ith DIP Sw itch
-20〜80
-20-80 °C or 0-170 °F
(humidity range 0-100%)
0 〜 50
0-50 °C or 30-130 °F
(humidity range 0-100%)
Manual setting