20
-
【▲/【▼
Chapter 6. Setting Recordeing Range
6-1. Setting temperature recording range
The temperature recording range
can be set by changing over the DIP
switches.
CAUTION
Before changing over the DIP
switch, be sure to turn OFF (
)
the power switch.
Setting range 0-50°C (30-130°F)
Turn OFF (
) the power switch, then turn ON the DIP
switch No. 3, and turn OFF the No. 4.
When the power switch is turned ON, temperature will be
recorded in the range from 0 to 50°C or 30 to 130°F.
Setting recording range manually
Turn OFF (
) the power switch, turn ON the DIP
switches No. 3 and No. 4, and then turn the power switch
ON again.
Press the
FUNCTION
key five times (Six times when
the SD card is inserted and the [SD Card Record] switch
is set to OFF) while the normal measurement screen is
displayed.
MODE
and
r1
will appear and the present
central temperature of the chart will blink on the left side.
- -
will blink instead if the recording range has not been
changed from the initial setting. (-20 to 80 °C or 0 to 170
°F)
Change the temperature to be recorded in the center of
the chart paper with the
【▲】
or
【▼】
key, then press the
ENTER
key. The central temperature will stop blinking
to show the setting has been completed.
Then, the present recording range will blink on the right
side of the screen.
Set the recording range 10 (+ / -10°C) or 25 (+ / -25°C)
in Celsius or 25 (+ / -25°F) or 50 (+ / -50°F) in Fahrenheit
with the
【▲】
or
【▼】
key, then press the
ENTER
key.
The recording range will stop blinking to show that set-
ting has been completed.
Reference
If this setting is not required, press the
FUNCTION
key twice to return to the normal display.
ON
3 4
ON
3 4
ON
3 4
Sett ing Temperature
Record ing Range w ith DIP Sw itch
-2080
-20-80 °C or 0-170 °F
(humidity range 0-100%)
0 50
0-50 °C or 30-130 °F
(humidity range 0-100%)
Manual setting