27
-
Press the
ALARM
key to return to the “Sensor calibra-
tion status” screen.
CAUTION
To cancel this setting, press the
ALARM
key and pro-
ceed to the next setting item.
10-3. Calibrating humidity sensor
Press the
ALARM
key twice while the “Sensor
calibration status” screen is displayed. The
1P
and
C3
indications will appear.
Press the
ENTER
key. The indication on the left
side will start blinking.
Use the
【▲】
or
【▼】
key to enter the humidity value
shown on the reference hygrometer, then press the
ENTER
key. The entered value will stop blinking to
show the setting has been completed.
Press the
ALARM
key to return to the “Sensor cali-
bration status” screen.
CAUTION
To cancel this setting, press the
ALARM
key and
proceed to the next setting item.
10-4. 2-point calibration of humidity sensor
Calibration at 33% RH
Place the sensor in a saturated salt type humidity cali-
brator or a reference humidity generating bath.
Press the
ALARM
key three times while the “Sensor
calibration statusscreen is displayed. The
2P
and
C5
indications will appear.
Press the
ENTER
key, and the present temperature
and humidity will be displayed (C6). Wait until the humid-
ity becomes stable in the range of 33% + / -5RH.
When the humidity is stabilized, press the
ENTER
key.
The
33.0
indication will start blinking.
Use the
【▲】
or
【▼】
key to enter the humidity value that
corresponds to the present temperature shown on the
humidity calibrator or the value indicated on the humidity
bath, then press the
ENTER
key.
The indication on the left side starts count from 1 to 5,
then the humidity indication will stop blinking to show the
setting has been completed.
Calibration at 75% RH
Place the sensor in a saturated salt type humidity cali-
brator or a reference humidity generating bath.
After the 33% RH calibration is completed, press the
ENTER
key. The
2P
and
C7
indications will ap-
pear.
Press the
ENTER
key, and
the present temperature
and humidity will be displayed (C8). Wait until the humid-
ity becomes stable in the range of 75% + / -5RH.