28
-
When the humidity is stabilized, press the
ENTER
key.
The
75.0
indication will start blinking.
Use the
【▲】
or
【▼】
key to enter the humidity value that
corresponds to the present temperature shown on the
humidity calibrator or the value indicated on the humidity
bath, then press the
ENTER
key.
The indication on the left side starts count from 1 to 5,
then the humidity indication will stop blinking to show the
setting has been completed.
Press the
ALARM
key to return to the “Sensor calibra-
tion status” screen.
CAUTION
When this setting has been completed, all the keys
but the
ALARM
key will be inactive.
If the 2-point calibration of the humidity sensor is
made, the humidity value calibrated in paragraph
10-3 will become invalid.
To cancel this setting, press the
ALARM
key and
proceed to the next setting item.
10-5. To initialize the calibrated value
Follow the procedure below to initialize the calibrated val-
ue. If it is not necessary, skip this paragraph.
When the “Sensor calibration status” screen (C0) is dis -
played, press the
ALARM
key 5 times. The
C9
indi-
cation will appear with the
CLEAR
indication blinking
on the screen.
Press the
【▼】
and
ENTER
keys together. The cali-
brated value will be cleared and the “Sensor calibration
status” screen (C0) will be restored.
CAUTION
To initialize means to restore the factory-setting.
To return to the normal measurement screen, turn
OFF (
) the power switch.
Chapter 11. Pen Position Adjustment
11-1. Function overview
Errors in recording due to extension or contraction of the
chart paper can be adjusted. If it is not necessary, skip this
chapter.
11-2. To check the adjustment status
Turn ON the power switch while pressing the
【▼】
and
ENTER
keys together. The “Pen position adjustment sta-
tus” screen will be displayed.
CAUTION
During this operation, no switch can be used until the
pens stop moving completely and the blinking indica-
tion becomes still.