METIS M308 / M309 / M313 / M316 / M318 / M323 / H309 / H316 / H318 (12-pin)
SensorTools Software
47
7
GG
Internal device temperature (in 1/256°C oder 1/256°F)
8
HH
Errror status (see 11, Interface commands: fs command)
9
- - - -
Unused
10+11
IIIIII
Device status (internal purpose)
13+14
JJJJ
Beginning of device temperature range (in 1/10 degree)
15+16
KKKK
End of device temperature range (in 1/10 degree)
17+18
LLLL
Beginning of temperature sub range (in 1/10 degree)
19+20
MMMM
End of temperature sub range (in 1/10 degree)
21-32
----
Unused
Mode 1 (Data packet size 64 Byte): Sequence and Meaning
Byte
Bytes
Meaning
1+2
AA
Measured temperature (in 1/10°C); special case: F000h = invalid measured value (overflow)
3+4
BB
Emissivity (in ‰ ) 1000 = 100%
5+6
CC
Clear settings of maximum value storage / Storage mode (0...4) 0 = No storage;
1 = time clear; 2 = Ext. clear; 3 = Automatic clear; 4 = pouring stream (optional)
7-10
DDDD
Resonse time (0...25 s); in steps of 10 µs (00000000h-0003d090h)
11-14
EEEE
Clear time (0...10 s); in steps of 10 µs (00000000h-000186a0h)
15+16
FF
Analog outputs (Byte 1= analog output 1, Byte 2=analog output 2)
Bit 7-Bit 4 = 020 mA, 420 mA; Bit 3-Bit 0= Output temperaturr, controller, etc.
17+18
GG
Internal device temperature (in steps of 1/256°)
19-22
HH
Errror status (see 11, Interface commands: fs command)
23+24
IIII
Status bytes
Bit Status byte 0 Status byte 1 Status byte 2 Status byte 3
Bit0 (lsb) Fahrenheit active Controlling active Setup0 Display 0
Bit1 Status dig. output 1 AutoTune active Setup1 Display 1
Bit2 Statusdig. output 2 AutoTune at controller start Setup2 Display 2
Bit3 Statusdig. output 3 Device ready Unused Unused
Bit4 Status dig. input 1 Device hardware error Unused Unused
Bit5 Status dig. input 2 Controller finished (successful) Unused Unused
Bit6 Status dig. input 3 Targeting light active Unused Unused
Bit7 (msb) Status dig. input 4 Unused Unused Unused
25+26
JJ
Beginning of device temperature range (in 1/10 degree)
27+28
KK
End of device temperature range (in 1/10 degree)
29+30
LL
Beginning of temperature range analog output 1 (in degree)
31+32
MM
End of temperature range analog output 1 (in degree)
33+34
NN
Beginning of temperature range analog output 2 (in degree)
35+36
OO
End of temperature range analog output 2 (in degree)
37+38
PP
Devices with controller: setpoint value (in 1/10 Grad)
39+40
QQ
Devices with controller: proportional band Xp (0…10000) in 0.1%
41+42
RR
Devices with controller: integral time Ti (0…65535) Factor × sampling time
34+44
SS
Devices with controller: derivative tme Td (0…65535) Factor × sampling time
45+46
TT
Devices with controller: minimum controller output power P-Min (0…999) in 0.1%
47+48
UU
Devices with controller: maximum controller output power P-Max (1…1000) in 0.1%
49+50
VV
Devices with controller: AutoTune target temperature (0…1000) in 0.1%
51+52
WW
Devices with controller: controller output power limitation P-dynamik (0…10000) in 0.01%
53+54
XX
Devices with controller: ramp time (0…65535) in 100 ms
55+56
YY
Devices with controller: hold time (0…65535) in 100 ms
57+58
ZZ
Devices with controller: sampling time (100...2500) in 10 µs
59+60
ÄÄ
Devices with motorized optics: smallest possible measuring distance (in mm)
61+62
ÖÖ
Devices with motorized optics: largest possible measuring distance (in mm)
63+64
ÜÜ
Devices with motorized optics: current set measuring distance (in mm)

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the SensorTherm METIS M3 series and is the answer not in the manual?

SensorTherm METIS M3 series Specifications

General IconGeneral
BrandSensorTherm
ModelMETIS M3 series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals