METIS M308 / M309 / M313 / M316 / M318 / M323 / H309 / H316 / H318 (12-pin)
SensorTools Software
47
Internal device temperature (in 1/256°C oder 1/256°F)
Errror status (see 11, Interface commands: fs command)
Device status (internal purpose)
Beginning of device temperature range (in 1/10 degree)
End of device temperature range (in 1/10 degree)
Beginning of temperature sub range (in 1/10 degree)
End of temperature sub range (in 1/10 degree)
Mode 1 (Data packet size 64 Byte): Sequence and Meaning
Measured temperature (in 1/10°C); special case: F000h = invalid measured value (overflow)
Emissivity (in ‰ ) 1000 = 100%
Clear settings of maximum value storage / Storage mode (0...4) 0 = No storage;
1 = time clear; 2 = Ext. clear; 3 = Automatic clear; 4 = pouring stream (optional)
Resonse time (0...25 s); in steps of 10 µs (00000000h-0003d090h)
Clear time (0...10 s); in steps of 10 µs (00000000h-000186a0h)
Analog outputs (Byte 1= analog output 1, Byte 2=analog output 2)
Bit 7-Bit 4 = 0–20 mA, 4–20 mA; Bit 3-Bit 0= Output temperaturr, controller, etc.
Internal device temperature (in steps of 1/256°)
Errror status (see 11, Interface commands: fs command)
Status bytes
Bit Status byte 0 Status byte 1 Status byte 2 Status byte 3
Bit0 (lsb) Fahrenheit active Controlling active Setup0 Display 0
Bit1 Status dig. output 1 AutoTune active Setup1 Display 1
Bit2 Statusdig. output 2 AutoTune at controller start Setup2 Display 2
Bit3 Statusdig. output 3 Device ready Unused Unused
Bit4 Status dig. input 1 Device hardware error Unused Unused
Bit5 Status dig. input 2 Controller finished (successful) Unused Unused
Bit6 Status dig. input 3 Targeting light active Unused Unused
Bit7 (msb) Status dig. input 4 Unused Unused Unused
Beginning of device temperature range (in 1/10 degree)
End of device temperature range (in 1/10 degree)
Beginning of temperature range analog output 1 (in degree)
End of temperature range analog output 1 (in degree)
Beginning of temperature range analog output 2 (in degree)
End of temperature range analog output 2 (in degree)
Devices with controller: setpoint value (in 1/10 Grad)
Devices with controller: proportional band Xp (0…10000) in 0.1%
Devices with controller: integral time Ti (0…65535) Factor × sampling time
Devices with controller: derivative tme Td (0…65535) Factor × sampling time
Devices with controller: minimum controller output power P-Min (0…999) in 0.1%
Devices with controller: maximum controller output power P-Max (1…1000) in 0.1%
Devices with controller: AutoTune target temperature (0…1000) in 0.1%
Devices with controller: controller output power limitation P-dynamik (0…10000) in 0.01%
Devices with controller: ramp time (0…65535) in 100 ms
Devices with controller: hold time (0…65535) in 100 ms
Devices with controller: sampling time (100...2500) in 10 µs
Devices with motorized optics: smallest possible measuring distance (in mm)
Devices with motorized optics: largest possible measuring distance (in mm)
Devices with motorized optics: current set measuring distance (in mm)