SCORPIO 38’ - 45’ - F.A.Q. / COMMON PROBLEMS - Possible problems related to the column
Version 1.02 May 27, 2022
SERVICEVISION BIS SL
Ríos Rosas, 20 · 08940 CORNELLA DE LLOBREGAT (Barcelona) Spain · Tel. 34 93 223 86 30 · Fax 34 93 223 86 31 81
comercial@servicevision.es · www.servicevision.es
7.3 POSSIBLE PROBLEMS RELATED TO THE COLUMN
The column does not move
manually
-Check if the screws of the column are loose.
The column does not move
with the Hand Command
-Check if the screws of the column are loose.
-Check if the column cable is connected.
-Change the column cable.
-Check the belt for the motor.
-Check if there is an alarm in the display.
-Contact Servicevision
7.4 POSSIBLE PROBLEMS RELATED TO THE COUNTERWEIGHTS
The counterweights make
noise in a point of the arm
-Check if the counterweight generator is activated and if so, check if
there are 2 points really close to each other with a high difference of
% on the counterweights.
The counterweights do not
move
-Check if the counterweights are enabled
-Check if the cable is connected.
-Contact Servicevision.
There is an alarm “CWE
LIMITS ALARM”
-Both counterweight limits are detecting at the same time.
-Check if there is something stuck on the counterweights trolley
-Check if the counterweight cable is properly connected.
-Contact Servicevision for further troubleshooting.