6.3 Selecng Gas Molecular Weights
The SDHmini default seng is for air (molecular weight 28.99), however, this
funcon allows the molecular weight to be set appropriately for other sample
gases, for example 18.99 .
6.4 Pressure
This sub-menu has, for both the primary and secondary displays, ve opons
allowing the user to select either atmospheric pressure or one of four pressure
engineering units and applied pressure.
When this funcon is selected, the main and secondary display screens will display
the pressure at which the dewpoint reading is calculated.
The SDHmini requires that a numerical pressure value be entered in the selected
units at the boom of the units selecon secon, for example selecng pressure
dewpoint of 2,000 psig on secondary display.
Warning: This funcon is purely an internal calculaon and DOES NOT allow an
elevated pressure to be applied to the measuring head.