EasyManua.ls Logo

Shaw SDHmini - Piping Schematic Component Index; Purging the Sample Connection

Shaw SDHmini
36 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
3.4 Piping Schemac Component Index
Sample Isolaon Valve - This is a recommended item as it allows access to
the sample system without interrupng the main process line.
Filter Unit – A lter unit is recommended when the samples are likely
to contain parculate maer. If the air/gas sample contains heavy
hydrocarbon condensate, the lter must be of the coalescing type with a
drain. The lter unit should be posioned as close to the sample point as
praccal.
Sample Tube – This should be stainless steel for dry air or gas applicaons
but copper or carbon steel can be used where weer gases are to be
measured. If any secon of the sample tube must be exible then PTFE
should be used. In most cases, 3 mm OD (⅛”) is sucient as it provides
good system response me with minimum ow. 6 mm OD (¼”) tube can
be used where pressure drops across the 3 mm tube are too high.
Pressure Reducon Valve or Pressure Regulator – the sample is measured
at atmospheric pressure requiring that valve 4 is ed to the system.
Flow Control Valve.
Sample Connecon.
SDHmini.
Flow Indicator – The recommended sample ow is 2 to 5 L/min.
Sample Exhaust – The exhaust is vented to atmosphere or returned to an
atmospheric pressure line.
3.5 Purging the Sample Connecon
Refer to the Piping Installaon Schemac in secon 3.3.
It is advisable to carry out an inial purge of the sample loop without connecng
the instrument, in order to avoid the possibility of sensor damage on start-up.
Open the sample isolaon valve (1) slowly, unl there is a small ow of air/gas at
atmospheric pressure from the pressure reducon valve (4). Allow the air/gas to
exhaust through the sample connecon (9).
Connue this for two minutes to remove any residual moisture from the sample
pipe work and components. Check that no liquid or parculate contaminaon,
which could damage the sensor, is passing through the sample pipe. Direcng the
sample onto a piece of white paper (such as a lter paper) will make it easier to
see any dust or oil mist in the sample.
1.
2.
3.
4.
5.
6.
7.
8.
9.
9

Table of Contents

Related product manuals