Chapter 6 Maintenance
6-12
AA-6800/6650 Instruction
Name P/N Remarks
Standard
Replacement
Interval
Lamp, D2 L6382 062-65063-01 Deuterium lamp 1 year
Nebulizer set 206-52300-91 With impact bead 2 years
Chamber ASSY 206-50363-91 Made of polypropylene 6 years
Burner head ASSY 206-50370-92 Made of titanium (10 cm) 3 years
Burner socket ASSY 206-50762-91 3 years
O ring, AS568-116-4D 036-15551-16 Used at the burner head side of the
burner socket
1 year
O ring, 4D P22 036-11222 Used at the chamber side of the
burner socket
1 year
O ring, 4D P5 036-11203 Used in the nebulizer 1 year
O ring, 4D P16 036-11217 Used with the nebulizer and chamber
plugs
1 year
O ring, 4D P14 036-11215 Used in the drain joint 1 year
O ring, 4D P35 036-11237 Used between the chamber and the
jacket
1 year
Cleaning wire 201-79229-01 For capillary cleaning
Sampling tube 204-05899-01 PTFE* tube for sample suction
Sampling tube ASSY 206-50772-91 Polyethylene tube for sample suction
Drain ASSY 206-80771-91 Overflow-type waste liquid trap 3 years
Gas tube ASSY 206-65112 Nylon tube set for gas supply 3 years
Hose, 8 u 2B-BK 016-02101-04 For air (black)
Specify the necessary length.
(Standard length: 5 m)
3 years
Hose, 8 u 2B-R 016-02101-03 For fuel gas (red)
Specify the necessary length.
(Standard length: 5 m)
3 years
Hose, 8 u 2B-G 016-02101-01 For N
2
O (green)
Specify the necessary length.
(Standard length: 5 m)
3 years
Wire band for 16 mm
diameter
037-61019 Hose clamp
6.5 Maintenance Parts