Chapter 2 Installation
2-10
AA-6800/6650 Instruction
Figure 2.4 Recommended Piping Setups for AA-6800/6650
2.3.4 Ventilation System
Place a ventilation duct with a hood above the atomizer.
(1) The hood, duct and ventilation fans used should be all made of metal. Those made of plastic are
inappropriate as they are softened by heat.
(2) A cooling-fan air capacity of about 600 to 1200 m
3
/h is appropriate. Too much suction force will cause the
flame to flicker as well as create excessive noise. Placement of a damper in the duct system is ideal for
setting the optimum air flow.
Figure 2.5 Typical Ventilation System
Drain
separator
Full scale 0.2 MPa
0.09 MPa
Full scale 0.2 MPa
0.09 MPa
Full scale 0.5 MPa
0.35 MPa
Full scale 0.5 MPa
0.35 MPa
Full scale 0.5 MPa
0.35 MPa
Stop cock
Analysis room
Stainless steel piping
(inside dia. 7 mm)
Stop cock
Acetylene
gas
Hydrogen
gas
Nitrous
oxide
gas
Argon gas
Air
compressor
Out-
doors
Damper
Fan
(Unit: mm)
About 700