2.4 Assembling the AA-6800/6650
2-23
AA-6800/6650 Instruction
2.4.6 Installation of PC Software
This procedure assumes that Windows is already installed on the PC. For its installation, refer to the
appropriate documentation.
(1) Start up Windows. In case of installing to the Windows XP, Windows 2000 or Windows NT4.0, logon to
Windows by Administrator authority.
(2) Insert CD-ROM for CD drive. Execute “\eng\setup.exe” of CD drive by using Windows explorer.
(3) First, a note for installation is displayed. Read it and click on <Next> to go ahead.
(4) Select AA-6800 or AA-6650 as main unit type that you use. Click on <Next> to go ahead.
(5) Select GFA-EX7 or GFA-6500 as GFA type that you use. Click on <Next> to go ahead.
(6) Check the installation destination folder. You may change the folder here, if desired. Click on <Next> to go
ahead.
(7) Select the program folder. Click on <Next> to start setup.
(8) Finally, when the message shows that the setup is completed, click on <Finish>.
When the [Yes, I want to read README.TXT] check box is check-marked, the README.TXT is displayed
by the Notepad. Be sure to read this, since the important information about AA software operation is
described in the README.TXT. Exit the Notepad, then installation procedure is finished.
1. Set all the electric power-saving functions to OFF both in the BIOS setup of PC and in the control panel
of Windows. Otherwise, the electric power-saving function may cause a trouble in communication
between the AA main unit and the PC.
How to display the BIOS setup screen is different depending on the PC model. Refer to the instruction
manual provided with the PC.
2. When upgrading the WizAArd ver. 1.xx to ver. 2.xx
Don't delete "wizaa.ini" file in the Windows directory when uninstalling the WizAArd ver. 1.xx. Upgrading
to ver. 2.xx is performed more smoothly when the "wizaa.ini" file of WizAArd ver. 1.xx remains. Also, It
is highly recommended for you to back up "wizaa.ini" file before uninstalling the older version WizAArd.
3. When upgrading the WizAArd ver. 2.1x to ver. 2.20
Don't uninstall the WizAArd ver. 2.1x when it is already installed.
If the WizAArd ver. 2.1x is uninstalled, "wizaa.ini." file and lamp history data (lamphist.ref file) are
deleted and, as the result, various parameters and its lamp use time are reset. For upgrading, execute
the setup disk of ver. 2.20 and install it to overwrite "ver. 2.1x" on the folder in which the old version
software has been installed.
To keep ver. 2.1x, install ver. 2.20 in another folder. At this time, to transfer the lamp history data from
ver. 2.1x, copy the "lamphist.ref" file of ver. 2.1x to the folder in which ver. 2.20 has been installed.
NOTE