EasyManua.ls Logo

Shimadzu AA-6800F - Page 40

Shimadzu AA-6800F
302 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Chapter 1 Overview
1-6
AA-6800/6650 Instruction
AA-6650F Specifications
AA-6650G Specifications
Burner Unit
Type Air-cooled pre-mix type
Burner Head Pure titanium
Nebulizer Pt-Ir capillary with PTFE orifice,
Impact bead is inseparable. Fluoric acid is usable.
Chamber Polypropylene
Positioning Manual positioning by adjusting knob
Gas Controller
Flow Rate/Pressure Control Automatic flow rate setting
Safety Measures Automatic gas leak check
•Air-C
2
H
2
flame priority ignition
•Air/N
2
O auto switching with C
2
H
2
flow rate monitor (optical
sensor)
Prevents raw gas discharge
Prevents usage of wrong burner head
Pressure monitor to prevent flashback at abnormal pressure
Safety extinction by automatic detection of momentary power
failure
Drain tank water level monitor to prevent flashback
Heating Temperature Range Ambient to 3000 qC
Heating Control Method Drying: Current control method
Ashing/Atomizing: Optical temperature control method
Heating Parameter Settings Max. 20 stages
Heating mode: RAMP/STEP
Inner gas type: 2-line switching
High sensitivity setting
Furnace concentration boost cycle: Max 20 times
Inner gas flow rate: 0 - 1.5 L/min
Optimum furnace program search function