5.2
Fundamentals
of
Operation
I
Note
I
I
Note
I
When the above operation is repeated, the mode is changed form
LA-+P+LJ
.
Mode
L
is only applicable when an optional
flow selection block is installed. When
the
block is
not
installed, do
not perform measurement in mode
L.
Refer
to
5.3.1 for the mode setting when using the flow selection
block.
Mode
I
Refractive
index
measurement range
I
Input step
0.01
step
for
0.01
to
1
1
step
for
1
to
500
1
A
1
0.Ol-500X1O4RIU
I
P-L
~1-smx1o"RIu
I
1
step
for
I
to 5000
Select the mode corresponding to the refractive index
of
the
sample
to
be
measured. When the index is
500
x
10"
FUU
or less, use
mode A.
Note that the setting of mode
P
has wider measurement range while
it
has
the larger baseline noise.
The range value will
default to 100 when
the
mode
is
changed
as
described below.
0
When
the
mode
is
switched
to
mode
P
or
L
during measurement
with the range
of
0.01
to
1
.OO
in mode
A.
0
When the mode is switched to mode
A
during measurement
with the range of 501
to
5000
in mode
P
or
L.
6.
Setting measuring range
The procedure to set the recorder range is
as
follows:
Changing the range
from
4
to
16
X
lo4
FUUFS
(1)
Press
[f..c)
once in the initial state to access the range
parameter (blinking).
balance
wr10-7
Rnpe
mode
(2)
Input
0,
@and
(E.t.f)
ba)uue
warn
rangs
mode
0
0
16
A
(3)
After
a
value is input, it
returns
to the initial
state.
5
-
8
-
RID
-
1OA