9.1
Component
Validation
2.
Checking the lamp intensity
Content
of
inspection
This
test checks whether the light source intensity
is
sufficient.
Test procedure
(1)
Fill the cell with distilled water.
(2)
Select the
[6.
MONITOR]
on the main menu
of
the SCL-IOA
to open the monitor screen.
(Example
of
MONITOR screen]
ACT.TMP
40.0
T0TAL.E
7566
(Lamp intensity)
BALANCE
10
LAh4P.m
86
____
DET.A
__________I______-________
......................................
(3)
Record the light intensity value.
AcceDtance criteria
TOTAL
EN
is
more
than
6,000.
3.
When the value
does
not fall within reference value,
first
clean the
flow cell. (Refer to the section 8.1 Cleaning the Flow Lines.) Then
increase the value of the LAMP VOLT to
6,000
or more. (Refer to
the section
5.5
Additional Functions.)
Checking the temperature control system
Content
of
inspection
This test verifies whether the temperature control is working
properly.
Test procedure
(1)
Select the
[3.
DET CONTROL] on the main menu
of
the
SCL-1OA
to
open the parameter control screen
of
the detector.
(2)
Record the value
of
CELL.T
as
the set temperature.
9
-
6
-
RID
-
1OA