8.2
Span
Adjustment
(Example)
When the measured refractive index is
102.4
R
x
old span factor is
0.90,
the new span factor is:
(1
1)
Input the new span factor obtained by the above formula into
SPAN
-
A.
(12)
Inject distilled water from the inlet port and finish the span
adjustment.
(13)
For
span adjustment of mode P/mode
L,
change the mode
setting in
(2)
above to
P
or
L,
and follow the procedures
from
(4)
above. Input the span value obtained at each mode
into
SPAN
-
P
and
SPAN
-
L
each.
[RIU]
and the
0.90
x
100/102.4
=
0.88
t
New
SPAN
-
A
8.2.2
When span adjustment
is
performed
using
the
LC
workstation
(1)
(2)
Turn
ON
the power
of
the
IUD-1OA
and
LC
workstation.
Set the parameters of the
RID-1OA
as
follows:
Example
1
(CLASS
-
VP)
'Ddecta
A
:
RID-IOA
MODE
171
RANGE
vj
WL
ELLT
AUXRNG
I
I
RIP1
OA
parameters
Q,
0
H
Example
2
(CLASS
-
LClO)
Eolariir
@GJ
0
-
Measure Mode
0
khairticai
0
1:Preparative
Cell
TernPAC1
140
I I
II
V
3:
1
.LIE
-
5
RIU/mV
I
d+!T-7
RnllmU
8
-
5