7.2
Control from the
SCL-1OA
Parameter
RANGE
RESP
(2)
Using function key
(
lf41
),
select the detector
where parameters are to be set. This is only necessary
when two detectors
(e.g.,
A
and
B)
are connected. When
a detector
B
is selected and only one detector is connected,
Description Setting range Minimum Unit
Defaultvalue
Range
for
recorder
in
mode
A
(X
RIU)
0,O.O
1-500
(0.01-1.00)
0.01
100
Range
for
recorder
in
modes
P
and
L
(X
RIU)
0,
1-5000
1
100
Response
1
-
10
1
5
(1-500)
1
I
DETECTOR
IS
NOT CONNECTED
1
is disulaved.
AUXRNG
(3)
Move the
cursor
to select the parameter with the cursor keys
(
It)
and
11)
),
input a desired value, and press
[s)
.
Set
and select the range and minimum parameter
steps after accessing the help line with the
[fu.c)
key.
To
set the parameters for two detectors, refer
to
item
(2)
above. Parameters which can be set in the
DET CONTROL
screen
are
listed in the table below:
(4)
AUXrange
1
-
4
1
2
W
Parameter summary
POL
1
Polarity
I I I
11
-
1
or
1
CELL.T
I
Cell
temmrature
setting
I
O(-OFF),30-60
I
0.1
I
401
MODE
I
Measurementmode
I
0
-
2
I
1
I
OI
(5)
The following operations are possible in the detector control
screen using the function keys.
I I
-
-
[=I
(I)
Outputs detector parameters being displayed to
the
printer of
the
Chromatopac (requires the current loop interface).
Resets the parameters (including parameters for other devices such
as
solvent delivery unit, etc.) in the file
No.
being displayed to the
initial values.
Copies parameter values
from
the
currently displayed file to another
file.
Performs zero adjustment
of
a detector
A.
-
RID
-
1OA