38
7. Calibration
Notes
• The weight symbol will blink for approximately two minutes before the
automatic span calibration occurs when a temperature change is detected.
If PSC starts while the balance is in use, press the [POWER] key to abort
that cycle.
• The following conditions must all be satisfied for the set time in order for the
PSC function to be executed. If these conditions are not satisfied within 1
minute of the specified time, automatic span calibration will not be executed
and that cycle will be skipped.
· The balance must be in mass display mode
· If in mass display mode, the display must be within a 30 count
· The balance must be in a stable state (the stability symbol must be
displayed if in the mass display mode)
· The load on the pan must be near zero
· No other automatic span calibration (PSC or timer CAL) must have
occurred during the same time
Note
Using a verified balance as a legal measuring instrument in using region:
When PSC, fully-automatic span calibration, is not activated, operator must
carry out span calibration with the built-in weight (refer to 7. 2.1) upon blinking
of the Weight Symbol.
If PSC starts while the balance is in use, press the [POWER] key to abort that cycle.