38
7. Calibration
Notes
The weight symbol will blink for approximately two minutes before the
automatic span calibration occurs when a temperature change is detected.
If PSC starts while the balance is in use, press the [POWER] key to abort
that cycle.
The following conditions must all be satisfied for the set time in order for the
PSC function to be executed. If these conditions are not satisfied within 1
minute of the specified time, automatic span calibration will not be executed
and that cycle will be skipped.
· The balance must be in mass display mode
· If in mass display mode, the display must be within a 30 count
· The balance must be in a stable state (the stability symbol must be
displayed if in the mass display mode)
· The load on the pan must be near zero
· No other automatic span calibration (PSC or timer CAL) must have
occurred during the same time
Note
Using a verified balance as a legal measuring instrument in using region:
When PSC, fully-automatic span calibration, is not activated, operator must
carry out span calibration with the built-in weight (refer to 7. 2.1) upon blinking
of the Weight Symbol.
If PSC starts while the balance is in use, press the [POWER] key to abort that cycle.