11.4
Outputting at Fixed Times (Interval Timer)
Not applicable to a verified balance as a legal measuring instrument in using region
Automatically outputs the displayed value at preset intervals. The “T” symbol (“T” of the Tare symbol) is
illuminated when the Interval Timer is activated.
1 Select menu item
50
and set the output interval
(00:01 = 1 sec to 99:59 = 99 minutes 59
seconds). Refer to 4.4 for numerical value input
methods.
2 In the interval timer standby state when the “T”
and the Stand-by symbols are both illuminated,
press the [PRINT] key.
The first data is output. Data will be automatically
output at the set time intervals thereafter.
3 To stop output, press the [POWER] key.
The balance returns to the interval timer standby
state in step 1).
Notes
• Use the [O/T] key to erase the tare or zero the balance at any time.
• Pressing the [POWER] key in the interval timer standby state brings the
power supply standby state.
Interval Timer setting can be cleared without menu operation, by pressing
the [POWER] key for about three seconds. This does not clear the set time
interval value.
To cancel from the menu, set the interval timer value on menu
50
to zero.
• Using the interval timer function to record data over a long period may
cause data error due to balance drift.
• Some instruments receiving the data may not operate normally if the set
time interval is short. To correct this, set the time interval to a longer
period.
• Not more than one of these seven functions can be set at a time. Setting
the second one with the menu will automatically clear the previously set
function; Auto Print(10.3), Auto Zero(10.4), Peak Hold(11.3), Interval
Timer(11.4), Add-on mode(11.5), Animal Weighing mode(11.6), Formulation
mode(11.7 ). If a function is set via the menu while another function is
already set, the original function will be automatically cancelled.