41
7. Calibration
2
4
1
3
5
6
7
11
13
12
8
9
10
14
Notes
“SEt” is displayed during the process. Leave the balance in a stable state
until the mass display appears as in step 5.
In PCAL, the value of the “weight to be loaded” cannot be changed.
Set the PCAL password using menu item
71
(The default value is 9999.)
7.3.5
PCAL Password Setting (UPX Series only)
Not applicable to a verified balance as a legal measuring instrument in using region
This password is necessary to access the PCAL function.
It is recommended that the balance administrator set this password to prevent an unauthorized person
from incorrectly calibrating the built-in calibration weight.
Select menu item
71
.
The numerical setting display appears. Enter a 4-digit number from “0000” to “9999”.
Refer to 4.4 for numerical input.
Note
When the menu is reset, the PCAL password is reset to “9999”.