These are applied measurement functions that can be used with the balance.
Only one of the functions in the menu group 4 (Refer to 4.3) (
41
to
53
) can be used at a time. When one
of the functions in menu group 4 is to be used with a weighing unit other than gram, select the function
from the gram-display first. Then, change to the other unit with the [UNIT] key. To release this function, use
the [POWER] key. For returning to the previous function, Last Menu Recall function is convenient.
There are some other combinations of functions that can be used together. Some functions employ the
value set in Zero Range setting (refer to 10.5).
11.1 Solid Specific Gravity Measurement
Solid specific gravity measurement refers to the measurement of the sample (solid) weight in the air and in
a liquid of known specific gravity (or density) and the calculation of the sample specific gravity (or density).
The d symbol is used to represent the solid specific gravity in this balance. The data output unit is DS.
Note
Use of the optional SMK101A, or SMK-102 Specific Gravity Measurement Kit
(refer to A-4.) is recommended for efficient measurements. When using the
Gravity Measurement Kit, refer to the instruction manual of the kit. Follow the
instruction below when a hanging pan and a tank are prepared by the user.
1 Select menu item
60-1
,
60-2
in advance.
60-1
: Select either the hold display mode or the
continuous display mode for the specific gravity
value display. Press the [CAL] key repeatedly
until [SG HoLd] appears.
Each time the [O/T] key is pressed the“
symbol on the left side of the display turns on or
turns off. The display is configured as shown
below according to the on/off state of
on : Hold display mode
off : Continuous display mode (the specific
gravity value is updated according to the change
of weight in water)
Press the [POWER] key repeatedly to return to
the mass display.
Press the [CAL] key to display the medium liquid
density setting screen.
60-2
: Enter the value of the medium liquid
density (or specific gravity (g/cm
3
)) (water,
alcohol etc.) in which the sample is immersed.
Refer to 4.4 for numerical value input.