65
11. Application Functions
2
4
1
3
5
6
7
11
13
12
8
9
10
14
11.3 Detecting Peak Values (Peak Hold)
Not applicable to a verified balance as a legal measuring instrument in using region
Detects the peak value of a fluctuating weight. “Peak value” is the highest or lowest value displayed in the
duration after it has changed beyond five times the Zero Range until the first stability detection after that.
The “P” symbol (“P” of Auto-Print symbol) is illuminated when the Peak Hold function is activated. Refer to
8.3 for stability detection details, and 10.5 for zero range details.
Select menu item
49
.
1 In the peak detection standby state with the “P”
symbol and the Stand-by symbols illuminated,
press the [O/T] key to tare the display.
2 Press the [PRINT] key. The Stand-by symbol
disappears and peak value detection starts.
3 Upon stability is detected, the operation of peak
value detection is ceased. The detected peak
value is displayed with the “P” symbol and
Asterisk simultaneously illuminated. The peak
value is output to external devices if connected.
After this first stability detection, the peak hold
function will not operate whether the load on the
pan changes or not.
4 Press the [POWER] key. The balance returns to
the peak detection standby state described in
the above step 1.