69
11. Application Functions
2
4
1
3
5
6
7
11
13
12
8
9
10
14
Notes
• When stability is detected and the displayed value is within Zero Range,
zeroing occurs automatically to maintain the zero display.
• When the [POWER] key is pressed in the Add-on standby state, the power
standby state is initiated.
• Add-on Mode setting can be cleared without menu operation, by pressing
the [POWER] key for about three seconds.
• Pretaring Value (10.7) function cannot be used with Add-on Mode.
• Refer to 8.3 for stability detection details, and 10.5 for zero range details.
• When add-on mode is on, unit switching and minimum display 1d/10d
switching are not allowed.
• When Add-on Mode is ON, fully-automatic span calibration by PSC (7. 3. 2)
or Clock-CAL (7.3.3) is not performed. Weight symbol keeps blinking
when span calibration is necessary. Span calibration with the built-in
calibration weight or external weights can be performed between sets of
measurements (Add-on standby).
• Not more than one of these seven functions can be set at a time. Setting
the second one with the menu will automatically clear the previously set
function; Auto Print(10.3), Auto Zero(10.4), Peak Hold(11.3), Interval
Timer(11.4), Add-on mode(11.5), Animal Weighing mode(11.6), Formulation
mode(11.7 ). If a function is set via the menu while another function is
already set, the original function will be automatically cancelled.
• Cannot be used together with the EP-100/-110 printer statistical
calculation function.