71
11. Application Functions
2
4
1
3
5
6
7
11
13
12
8
9
10
14
(Operational condition setting check display)
When Cond 3 is selected
When Cond 2 is selected
When Cond 1 is selected
(Display)
Notes
• The pouring mode environmental settings will not be displayed even when
using pouring mode (8.2).
• The animal weighing mode operation mode setting verification will be
given priority.
(Operation)
The Animal symbol is illuminated when this mode is active.
In addition, the ◀ symbol which indicates
the operation mode setting (refer to the "Operation mode switching" item) will be displayed on the right of
the display.
1 Load the weighing vessel and press the [O/T]
key to zero the display.
2 Load the sample (animal) with a mass more than
50 times the Zero Range.
3 When the measurement reaches stability criteria
of Animal Weighing, the weighed value is
automatically output.
4 Unload the animal or press the [O/T] key
5 When the mass display becomes less than 10
times the Zero Range and stability is detected,
automatic zeroing occurs. The mass of any
residue remaining on the balance (excrement or
fur) is canceled and returns to zero display.