34 MCS300P · Technical Information · 8013265 V1-2 · © SICK AG
Operation
Subject to change without notice
Filtered values (FVi)
Menu: MCS300P/Parameterization/Variables and Functions/Filtered values
This menu shows the configured filtered values.
Designation Remark
Index Consecutive number of the filtered value (FV01, FV02, ...).
Source Tag.
Filter Type Average:
Arithmetic mean.
Suitable for noise-corrupted data without exceptional values.
Median:
Values are sorted according to size and the mean value is taken. Suitable
for data that can contain exceptional values.
1st derivation
2nd derivation
Averaging Period Unit: Seconds.
Interrupt. Source Source of interruption (IS) from tags.
Effective for "Average" and "Median": Filtering is suspended, the unfiltered
signal is output.
IS inverted Checkmark: Interruption source inverted.
IS holds Checkmark: Hold the filter output for interruption source.