5.2. Setting the Station Address ULTRAMAT 23
Note
The basic principles of operation are explained in chapter 5 of the ULTRAMAT 23
device manual.
The station address can be set in the main menu under Configuration with the Special Functions submenu
item. The level 2 code is required for this.
7 Warning
Changing the station address can affect communication on the bus.
Codes/Language
AUTOCAL-drift
ELAN / PROFIBUS
Fact.Data/Res/Units
Select the ELAN/PROFIBUS subitem
from this menu
Next select
PROFIBUS Parameters subitem
ELAN Parameter
ELAN ext.Interfer.
PROFIBUS Parameters
The PROFIBUS station address can be
set in this dialog.
Addresses between 0...126 can be set.
PROFIBUS
Address : 126