2.8.2 Installing the Optional ULTRAMAT / OXYMAT / CALOMAT 6 Field Housing Module
Note
The A5E00015899 Exi module for intrinsically safe connection to the PROFIBUS-PA may only be
installed in Exp[ia/ib] field housing devices.
With all other non-Ex devices only the A5E00034504 for non-intrinsically safe connection to
PROFIBUS-PA or the A5E00019145 module for connection to the PROFIBUS-DP can be installed.
Installation instructions:
Disconnect the device from the mains
Open the left-hand housing door
Remove the sheet metal cover
Remove the sheet metal cassette from the device
Remove the old optional module if installed
Fit the module in the rail
Fasten the module with the M3 screws to the underside of the cassette between connectors
Fit the bracket on the top edge of the optional module
Connect the connectors to the motherboard
Fit the sheet metal cassette in the device
Connect the 37 Sub-D-connector with the terminal block
Replace sheet metal cover
Connect the device to the PROFIBUS
Connect the device to the mains
Cassette
Motherboard
Connect 26-pin
ribbon cable
Optional module
Fig. 2.2