EasyManua.ls Logo

Siemens CALOMAT 6 - Transducer Block with ULTRAMAT, OXYMAT and CALOMAT

Siemens CALOMAT 6
140 pages
Go to English
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
ULTRAMAT 6 / OXYMAT 6 / CALOMAT 6 / ULTRAMAT 23 / OXYMAT 61
Schnittstelle PROFIBUS DP / PA Interface
A5E00054148-03
92
On ULTRAMAT 23 there are the following possibilities due to the O
2
measuring components (sensor):
Variants IR measuring
component 1
IR measuring
component 2
IR measuring
component 3
O
2
(if applicable)
7MB 2331 TB/AI 1 ---------------------- ---------------------- TB/AI 2
7MB 2333 TB/AI 1 TB/AI 2 ---------------------- TB/AI 3
7MB 2334 TB/AI 1 TB/AI 2 TB/AI 3 TB/AI 4
The corresponding gas for the measuring components is indicated by the device MLFB number (see rating
plate).
3.1.1 Transducer Block with ULTRAMAT, OXYMAT and CALOMAT
The analyzer transducer block receives its input value from the ULTRAMAT, OXYMAT and CALOMAT basic
unit. This causes the device-related functions of sensors as well as the measured-value acquisition and
processing to be separate from the function block.
Rel.
Index
Parameter Description
8 COMPONENT_NAME
Name of the component
9 PV Measured value which is forwarded to the AI block
10 PV_UNIT Measured value unit
11 PV_UNIT_TEXT Unit in text form
12 ACTIVE RANGE
Current measuring range/always "1". ULTRAMAT, OXYMAT
and CALOMAT use the largest measuring range here
(characteristic curve upper value)
13 AUTORANGE_ON Automatic measuring range switchover TRUE/FALSE
No significance as ULTRAMAT, OXYMAT and CALOMAT
always use the largest measuring range (characteristic curve
upper value).
14 SAMPLING_RATE Time between two measured value readings
25 NUMBER_OF_RANGES Number of measuring ranges/always just one measuring range
26 RANGE_1 Measuring range lower and upper value (“largest measuring
range“)
34 POSITIONING_VALUE
Value (1 byte) for triggering autocalibration. Taken from DO
block (OUT_D).
35 FEEDBACK Value (1 byte) for feedback on the autocalibration status (taken
from DO block FEEDBACK).