7.4.2 CALOMAT 6 Diagnostic Information
Fault (S)
Maintenance
request (W)
No.
Meaning of fault/maintenance request Diagnostic
byte/bit
Diagnosis bit name
S1 Parameter memory 1 / 4
2 / 2
DIA_MEM_CHKSUM
DIA_CONF_INVAL
S4 External fault 1 / 1 DIA_HW_MECH
S5 TCD temperature 1 / 1 DIA_HW_MECH
S12 Power supply 2 / 1 DIA_SUPPLY
S14 Measured value > characteristic curve
upper value
1 / 1
1 / 5
DIA_HW_MECH
DIA_MEASUREMENT
S15 Calibration cancelled 1 / 5 DIA_MEASUREMENT
W1 Calibration difference too large 2 / 5 DIA_MAINTENANCE
W4 Set clock 2 / 5 DIA_MAINTENANCE
W6 Temperature sensor LCD display 1 / 3
2 / 5
DIA_TEMP_ELECTR
DIA_MAINTENANCE
W9 External maintenance request 2 / 5 DIA_MAINTENANCE
W10 Autocal check difference 2 / 5 DIA_MAINTENANCE