ULTRAMAT 6 / OXYMAT 6 / CALOMAT 6 / ULTRAMAT 23 / OXYMAT 61
Schnittstelle PROFIBUS DP / PA Interface
A5E00054148-03
93
3.1.2 Analog Input Block with ULTRAMAT, OXYMAT and CALOMAT
The measured value and status for the process is provided via the analog input block.
Rel.
Index
Parameter Description
10 OUT Output value and status. Result of the analog input function block
11 PV_SCALE Input scale. As a rule the measuring range of the transducer block. The
lowest value (0%) is represented by 0, the top value (100%) by 1.
12 OUT_SCALE Output scale. Representation of user-specific physical size. The lowest
value (0%) is represented by 0, the top value (100 %) by 1. Indication of
user-specific unit.
13 LIN_TYPE Selection of curve shape for characteristic curve transformation.
16 PV_FTIME
Filter time constant τ of digital filter 1st order.
17 FSAFE_ TYPE Parameterization of safe position. Active only in AUTO mode if the
measured or simulation value has a bad status (BAD) or the current mode
O/S is active.
• 0: The default value FSAFE_VALUE is used for the safe position. The
status raised to uncertain – substitute set.
• 1: The output value last valid before the failure will still be used. The
status is raised to uncertain – last usable value. If there was no valid
value before, the initial value will be returned with the uncertain – initial
value status.
• 2: The value will be used for further calculation despite the bad status.
If this parameter is not implemented, the AI will behave as described in • 1.
18 FSAFE_VALUE Default value for safe position.
19 ALARM_HYS
Limit value hysteresis. For avoiding an alarm surge or waving of signal
flags. After a limit value has been exceeded the value must have moved
the hysteresis in the opposite direction before the limit value signal is reset.
21 HI_HI_LIM Upper alarm limit
23 HI_LIM Upper prewarning limit
25 LO_LIM Lower prewarning limit
27 LO_LO_LIM Lower alarm limit
30 HI_HI_ALM
Alarm status parameter for upper alarm limit.
31 HI_ALM
Alarm status parameter for upper prewarning limit.
32 LO_ALM Alarm status parameter for lower prewarning limit.
33 LO_LO_ALM Alarm status parameter for lower alarm limit.
34 SIMULATE Simulation value and status, enabling of simulation.
35 OUT_UNIT_TEXT ASCII text to be stored if the unit “text reference“ is indicated in
OUT_SCALE.