ULTRAMAT 6 / OXYMAT 6 / CALOMAT 6 / ULTRAMAT 23 / OXYMAT 61
Schnittstelle PROFIBUS DP / PA Interface
A5E00054148-03
132
Function Block AI 3 (Analog Input) cont.
Class Sl
ot
Index
abs
Ind
ex
rel
Elem
ent
Default
Value
Object Parameter description Data Type Num
ber
Store
(acc. to
profile)
cy
cli
c
re
ad
wri
te
MAND 33 17 0 FSAVE_TYPE Output / failure behavoir: type of reaction in
case of sensor failure
unsigned8 1 S x x
MAND 34 18 0 FSAVE_VALUE Output / safety default value in case of power
failure
float 1 S x x
MAND 35 19 5.0 ALARM_HYS Output / safety default value in case of power
failure
float 1 S x x
MAND 37 21 100.0 HI_HI_LIM Output / Output limit values / High alarm limit float 1 S x x
MAND 39 23 100.0 HI_LIM Output / Output limit values / High warning limit float 1 S x x
MAND 41 25 0.0 LO_LIM Output / Output limit values / Low warning limit float 1 S x x
MAND 43 27 0.0 LO_LO_LIM Output / Output limit values / Low alarm limit float 1 S x x
MAND 46 30 HI_HI_ALM High alarm limit value status ds39 1 D x
1 Unacknowledged Not acknowledged
2 Alarm State Alarm status
3 Time stamp Time stamp
4 Subcode Subcode
5 Value Value
MAND 47 31 HI_ALM High warning limit value status ds39 1 D x
1 Unacknowledged Not acknowledged
2 Alarm State Alarm status
3 Time stamp Time stamp
4 Subcode Subcode
5 Value Value
MAND 48 32 LO_ALM Low warning limit value status ds39 1 D x
1 Unacknowledged Not acknowledged
2 Alarm State Alarm status
3 Time stamp Time stamp
4 Subcode Subcode
5 Value Value
MAND 49 33 LO_LO_ALM Low alarm limit value status ds39 1 D x
1 Unacknowledged Not acknowledged
2 Alarm State Alarm status
3 Time stamp Time stamp
4 Subcode Subcode
5 Value Value
MAND 50 34 SIMULATE Output / Simulation: simulation of the
measured value (PRIMARY_VALUE in
Transducer Block
ds50 1 S x x
1 0x4b Simulate-Status Output / Simulation: simulation status
2 0.0 Simulate_value Output / Simulation: simulation value
3 0 Simulate_En/Disable Output / Simulation: simulation enable –
disable. Default disable (0)
OPT 51 35 OUT_UNIT_TEXT Unit text, valid, if OUT_SCALE.UNIT set to
"textual unit definition" (1995)
visible_string 16 S x x
52 36 bis reserviert von PNO
60 44
VIEW 65 49 VIEW_1 View-Objekt 1 AI void 4 D x
1ST_REV
6MODE_BLK
7 ALARM_SUM
10 OUT