ULTRAMAT 6 / OXYMAT 6 / CALOMAT 6 / ULTRAMAT 23 / OXYMAT 61
Schnittstelle PROFIBUS DP / PA Interface
A5E00054148-03
95
Status coding for ULTRAMAT/OXYMAT/CALOMAT:
Status coding for "bad quality"
Dec Hex Cause Remedy
11 0B Component missing Change configuration, install analyser including this
component
12 0C Device error Check with SIMATIC PDM / change analyzer
31 1F Out of order (equivalent value,
last valid value, through connect
value)
Depending on the paramterization. Switch in mode AUTO
Status coding for "uncertain quality"
Dec Hex Cause Remedy
64 40 Function control Warm up, AUTOCAL, in situ service
71 47 Measured value is bad,
last valid value
Check with SIMATIC PDM / change analyzer
75 4B Measured value is bad,
equivalent value
Check with SIMATIC PDM / change analyzer
79 4F After switching on the power,
initial value
Reject the measured value
80 50 Measured value is out of range Check with SIMATIC PDM / ambient temperature too high
Status coding for "good quality"
Dec Hex Cause Remedy
128 80 o.k., normal operation You can evaluate the measured values
132 84 Active block alarm, parameter
has been changed
Disappears after 10 seconds
137 89 Lower warning limit exceeded user programm
138 8A Upper warning limit exceeded user programm
141 8D Lower alarm limit exceeded user programm
142 8E Upper alarm limit exceeded user programm
164 A4 Maintenance request Check with SIMATIC PDM