5 - 1
Siemens AG SPB7-230.814.01 Page 1 of 24 Opdima
Medical Solutions Rev. 01 10.04 CS PS 24 Installation
Installation 5
General 5
Before starting the installation, a backup of the installation parameters of the
MAMMOMAT Novation
DR
can be carried out. This may be omitted, if backup already
available.
1. Connect a PC to the MAMMOMAT Novation
DR
and start the service program.
2. Select
Main menu => Backup => Copy installation area to disk => All
.
This will store the installation parameters on the hard disk of the PC.
3. Select
Main menu => Backup => Copy installation area to floppy => All
.
This will store the installation parameters on the service diskette.
4. Switch off the MAMMOMAT.
Sub-assemblies, overview 5
The components shall be placed according to the figure below.
The workstation shall be placed at least 1.5 m from the object
table of the stand.
X
Y
Z
mm
mm
mm
X
Y
Z
X
Y
Z
Z
X
Y
OPD00597
Biopsy
controller
Biopsy unit
CCD camera
Generator
Workstation
Stand with
X-ray unit
Workstation table
Fig. 1 MAMMOMAT 3000 Nova with digital biopsy and spot imaging system Opdimar
NOTE