Installation 5 - 21
Siemens AG SPB7-230.814.01 Page 21 of 24 Opdima
Medical Solutions Rev. 01 10.04 CS PS 24 Installation
Others:
select Frequency phase,
Frequency: 64*
Phase 160*
* Can be adjusted for optimal image reproduction, see the monitor manual “High-Resolu-
tion 18” LCD Monitor SCD 1898-0” (delivered with the monitor). Use the SMPTE test pat-
tern image for fine adjustments of the image.
select Sharpness, 3
select Video/PIP settings, off
select PIP Source, auto
select OSD
Horizontal: Max
Vertical: Max
Background: Translucent
Timeout: 20 Sec.
Make a final evaluation of the monitor settings with the clinical image as reference. See
Quality Control Manual.
Connecting the workstation to the network 5
The Opdima workstation can be connected to a network, partly for the use of DICOM and
also for networked printers.
1. Connect one end of the network cable (included in the delivery) to the workstation.
2. Connect the other end of the network cable to a network socket.
3. Log in with the user name “service”. The password is obtained from Siemens
Service Centre.
4. Press the Service button in the mode selection dialog, see Selection of mode in
MAMMOMAT 1000/3000/3000 Nova - Opdimar Service Instructions.
5. Press the Advanced service button in the service dialog, see Use of Advanced
service functions in MAMMOMAT 1000/3000/3000 Nova - Opdimar Service
Instructions.
6. In Advanced service configure the network parameters, see Network settings in
MAMMOMAT 1000/3000/3000 Nova - Opdimar Service Instructions.
For physical limitations of the network see Mammomat 1000/3000/3000 Nova -
Opdimar Planning Guide.