Opdima SPB7-230.814.01 Page 2 of 4 Siemens AG
Installation Rev. 01 10.04 CS PS 24 Medical Solutions
6 - 2 Final procedures
Performing of DHHS maintenance, see Page 6 - 3
Starting up the MAMMOMAT 6
Switch on the MAMMOMAT.
Starting up the workstation and login 6
The workstation is started according to instructions in Starting up the workstation and
login in Supplement to the Instructions for Use MAMMOMAT Novation
DR
- Opdimar.
Use the user name “service” and the password obtained from Siemens service center
when entering the system.
To obtain optimal picture, adjust the monitor settings as follow:
Brightness: 50
Contrast: 50
Backlight: 70
Starting up the MO unit 6
1. Insert the MO disk, type 9.1 GB MO disk
2. For switch settings and SCSI ID, see Paragraph “ Installing the MO unit” in
Chapter “Removal and replacement of sub-assemblies” in
MAMMOMAT 1000/3000 Nova - Opdimar Service Instructions.
Installation of software on OPDIMA WORKSTATION 6
The software is installed in the factory before delivery. If the software for any reason needs
to be reinstalled, refer to chapter “Reinstallation of software” in MAMMOMAT 1000/3000
Nova - Opdimar Service Instructions.
The Opdimar external diaphragm must be used for tests and cali-
brations involving radiation.
Any adjustments of the monitor settings shall be performed by
qualified personell who are employed by Siemens or one of its
affiliates or who are otherwise autorized by Siemens or one of its
affiliates to provide such services.
NOTICE
NOTE
OPD00601