Opdima SPB7-230.814.01 Page 8 of 24 Siemens AG
Installation Rev. 01 10.04 CS PS 24 Medical Solutions
5 - 8 Installation
Installation of cable duct 5
Four cables are laid from the bottom of the biopsy controller in a cable duct to the cable
outlet cover (3/Fig. 8). To install the cable duct on the rear side of the stand proceed as
follows:
1. Attach the cable duct (2/Fig. 8) to the rear cover by removing the protective strips
from the attached Velcro strips and place the duct flush with the biopsy controller
(1/Fig. 8). Secure the cable duct to the two standoffs (5/Fig. 8) with the delivered
screws and washers.
Modification of cable outlet cover 5
To allow the cables to pass through the cable outlet cover, the cover has to be modified
according to the following:
1. Cut an opening centrally on the cable outlet cover (4/Fig. 8). The opening should
be approximately 50 x 25 mm.
2. Attach the supplied plastic strip around the opening of the cable outlet cover
(4/Fig. 8).
3. Install the cable outlet cover (4/Fig. 8).
4. Mount the previously removed stand covers.
UP
OPD00164
4
3
1
2
50mm
25mm
5
Fig. 8 Cable duct